| Insulation resistance | 
								
									| Test voltage | 100V | 250V | 500V | 1000V | 
								
									| Measuring range (Auto range)
 | 4.000/40.00/ 200.0MΩ
 | 4.000/40.00/400.0/2000MΩ | 
								
									| First Effective Measuring range | 0.1~20MΩ | 0.1~40MΩ | 0.1~200MΩ | 0.1~1000MΩ | 
								
									| Mid-scale value | 5MΩ | 50MΩ | 
								
									| Accuracy | |±2%rdg±6dgt | 
								
									| Second effective measuring range lower | 0~0.099MΩ | 
								
									| Second effective measuring range upper | 20.01~ 200.0MΩ
 | 40.01~ 2000MΩ
 | 200.1~ 2000MΩ
 | 1001~ 2000MΩ
 | 
								
									| Accuracy | |±5%rdg±6dgt | 
								
									| Rated current | DC 1~1.2mA | 
								
									| Output short circuit current | 1.5mA max | 
								
									| Ω/Continuity | 
								
									| Auto range | 40.00/400.0Ω | 
								
									| Accuracy | |±2%rdg±8dgt | 
								
									| Output Voltage on open circuit | 5V±20% | 
								
									| Output short circuit current | DC 220±20mA | 
								
									| Fuse | Quick acting ceramic fuse 0.5A/600V (Ø6.35×32mm) | 
								
									| AC Voltage | 
								
									| Range | AC 20~600V (50/60Hz) DC -20~-600V/+20~+600V
 | 
								
									| Accuracy | |±3%rdg±6dgt | 
								
									| General | 
								
									| Applicable standards | IEC 61010-1 CAT.III 600V, IEC 61557-1,2,4, IEC 61326-1 (EMC), IEC 60529 (IP40) | 
								
									| Dimensions / Weight | 105(L) × 158(W) × 70(D)mm / 600g approx. | 
								
									| Power source | R6P×6 or LR6×6 | 
								
									| Accessories | 7150A (Remote control switch set), 9121 (Shoulder strap), R6P (AA)×6, Instruction Manual
 | 
								
									| Optional | 7115 (Extension Probe), 8016 (Hook type prod), 8923 (Fuse [0.5A/600V]), 9089 (Carrying Case)
 |