DT-570-TA |
SOLDERING IRON CHECKER (with calibration certificate) |
|
|
|
|
* |
Because the certificate will bear the purchaser's name (company name, etc.) , your name is required when ordering. |
|
|
|
DT-570 measurement method |
(1) |
Turn the POWER ON, and set the MODE to Ω or mV. Connect the dedicated ground cord to the unit, then touch one end to the plate and record the measurement (V1 or R1) (See Fig.1). |
(2) |
Using the dedicated adapter, connect the 3-wire type soldering iron, then turn the power ON and set to the maximum temperature (see Fig.2). |
(3) |
Connect the dedicated ground cord and the dedicated adapter.
After the temperature stabilizes, touch the tip in the center of the plate, then record that value (V2 or R2).
Use the recorded values and the formula below to make the calculation.
Leakage voltage V= V2 - V1, or Ground resistance R = R2 - R1
Example: Leakage voltage
When V1=0.1 and V2=0.7, V = 0.7 - 0.1 = 0.6 mV. |
|
|
|
|
|
|